Caractéristiques techniques :
SP1-CX | MSP1-CX | |
Main Spec | ||
Accuracy | ≤1.0 % (syringe ≥1 mL) | |
Precision | CV≤0.05% at full stroke (with syringe ≥ 500uL) | CV≤0.05% at full stroke (with syringe ≥1mL ) |
Wetted material | Borosilicate glass, PTFE, PFA, CTFE, ETFE, UHMW-PE,Ceramic | |
Dimension(L×W×H) | 114mm x 65mm x 254mm | 110mm x 44.4mm x127mm |
Weight | 2.15 kg | 1.0 kg |
Power supply | DC 24V, 1.5A (peak) | |
Syringe and Drive | ||
Syringe fitting | 1/4″-28 thread | |
Syringe | 50uL, 100uL, 250uL, 500uL, 1mL, 2.5mL, 5mL, 10mL, 25mL | 50uL, 100uL, 250uL, 500uL, 1mL, 2.5mL, 5mL |
Syringe material | Borosilicate glass, stainless steel plunger with PTFE coating, PTFE or UHMW-PE plunger tip | |
Resolution (standard mode) | 6000 steps with 0.01mm in 1 step | 3000 steps with 0.01mm in 1 step |
Resolution (microstep mode) | 48000 microsteps with 1.25um in 1 microstep | 48000 microsteps with 0.625um in 1 microstep |
Travel | 60mm | 30mm |
Time for full stroke | 1.26s-1200s | |
Drive system | Ball screw drive with optical encoder | |
Valve & Valve Drive | ||
Turn Time | ≤250ms between adjacent ports | |
Valve options | 3-port Y valve, 3-port distribution valve, 4-port valve, 6-port distribution valve, 9-port distribution valve, T valve, straight-through valve or distribution valve. | |
Max pressure | 0.68MPA | |
Fitting | 1/4″-28 thread | |
Communication | ||
Interface | RS485, RS232, CAN bus, external control input and output | |
Protocol | Data terminal, OEM protocol (serial) or CAN | |
Baud rate | RS485 or RS232: 9600bps or 38400bps CAN bus: 100 KB, 125 KB 250 KB, 500 KB or 1MB | |
Format | Data bit: 8; parity: none; stop bit: 1; half duplex (RS232/485) | |
Firmware | ||
Topology group | Up to 15 pumps in one group, support group control and individually addressed | |
Upgrade firmware online | "One Click" by customer through PC Utility | |
Homing Algorithm | Plunger home position is identified by motor stall, which will reduce the dead volume of the system | |
Halt function | For better interactive control, "h" command is used to halt execution of the command string during plunger movement. To resume execution with "R" or "r" command | |
Microstep mode | Set microstep mode for smoother motion with 48000 microsteps per stroke, 1/8 microstepping or 1/16 microstepping | |
Programmable parameter | Programmable ramps, cut-off velocity, plunger speeds, backlash compensation, programmable work-flow command set , delays and loops, terminate moves, diagnostics, absolute and relative positions |
|
Environmental | ||
Operating temperature | Recommended: 15℃-40℃ , use low temperature syringe for low temperature application | |
Operating Humidity | <80% RH, no condensation | |
Storage temperature | -20℃-65℃ | |
Storage Humidity | <80% RH, no condensation |