Specifications Crystal Frequency Mass Sensitivity (Hz/gm/cm2) Dynamic Range (gm) 3 MHz 10 MHz 15 MHz 2.04 x 107 (Hz/gm)cm2 2.26 x 108 (Hz/gm)cm2 5.09 x 108 (Hz/gm)cm2 4.66 x 10-4 gm   1.40 x 10-4 gm   9.33 x 10-5 gm  Dimensions Without Heat Sink (in.) With Heat Sink (in.)...

Description

Specifications
Crystal Frequency Mass Sensitivity
(Hz/gm/cm
2)
Dynamic Range
(gm)
3 MHz
10 MHz
15 MHz
2.04 x 107 (Hz/gm)cm2
2.26 x 108 (Hz/gm)cm2
5.09 x 108 (Hz/gm)cm2
4.66 x 10-4 gm  
1.40 x 10-4 gm  
9.33 x 10-5 gm 
Dimensions Without Heat Sink (in.) With Heat Sink (in.)
Mark 16
Mark 17
Mark 18
Mark 19
Ø 0.855 x 0.830
Ø 0.855 x 0.995
Ø 0.855 x 1.175
Ø 0.855 x 0.633
Ø 0.855 x 0.855 x 0.125
Ø 0.855 x 0.855 x 0.125
Ø 1.350 x 0.225 x 0.125
Ø 0.855 x 0.855 x 0.125
Temp. Version Standard Temperature Low Temperature
Supply Voltage 8 to 12 VDC
(10 VDC Nominal)
8 to 12 VDC
(10 VDC Nominal)
Power
In-Situ Heater
~120 mW @ 10 VDC
<5.8 W (MK16, 17, 18)
2.45 mW @ 10 VDC
Signal Amplitude >6 to 10 Vp-p 0.2 to 2.0 Vp-p
Voltage Sensitivity <23 Hz/V <23 Hz/V
Output Impedance 12K OHMS 13K to 27K OHMS
Temperature Range -199 to 100°C 10 to 400°K
Weight Grams
Mark 16
Mark 17
Mark 18
Mark 19
28.5 gm
19.7 gm
28.9 gm
19.7 gm